Power Requirements
IPm=Peak Module Current
IDm=Dynamic Module Current
Power Available for SCPs
(See the VXI Catalog or SCP
User's Manual for SCP current)
Cooling Requirements
Measurement ranges
DC Volts
Temperature
Resistance
Strain
Measurement resolution
Maximum
[SENSe:]FILT OFF
reading rate
[SENSe:]FILT ON
Appendix A
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Except where noted, all references to the HP E1413 apply to the HP E1313.
These specifications reflect the performance of the HP E1413 with the
HP E1413 Option 11 Straight-Through Signal Conditioning Plug-on. The
performance of the HP E1413 with other SCPs is found in the Specifications
section of that SCP's manual.
+5V
+12V
IP
ID
IP
IDm
m
m
m
0.20
0.
0.06 0.01 0.01 0.01
HP E1313A:
HP E1413C:
Average Watts/Slot
34
(Opt 11 or 12) ±62.5mV to ±16V Full Scale
Thermocouples - -200 to +1700 °C
Thermistors - (Opt 15 required) -80 to +160 °C
RTD's - (Opt 15 required) -200 to +850 °C
(Opt 15 with opt 11) 512 ohms to 131 Kohms FS
25,000 µe or limit of linear range of strain gage
100K rdgs/sec divided by the number of channels in the scan - for example,
100K rdgs/sec ÷ 64 channels = 1.56K rdgs/sec/ch
100K rdgs/sec ÷ 16 channels = 6.25K rdgs/sec/ch
6.4K rdgs/sec divided by the number of channels in the scan - for example,
6.4K rdgs/sec ÷ 64 channels = 100 rdgs/sec/ch
6.4K rdgs/sec ÷ 16 channels = 400 rdgs/sec/ch
Appendix A
Specifications
-12V
+24V
IP
ID
IP
ID
IP
m
m
m
m
0.5
0.1
0.5
.55A ±24V, 3.5A 5V
1A ±24V, 3.5A 5V
∆Pressure (mmH
O)
2
0.08
16 bits (including sign)
Specifications
-24V
-5.2V
ID
IP
ID
m
m
m
m
0.1
0.4
0.1
Air Flow (liters/s)
0.08
293

Advertisement