The registers used are:
Status Register (base+04
Parameter Register(s) (base+0A
Command Register (base+08
Scan Status and Control Register (base+10
The parameters must be written to their registers before the
command is written. Execution begins as soon as the command is
written.
If the command returns data, query for it before executing the
command.
ERRFLAGS?
Some of the register-based commands return information kept by the
on-board microprocessor. When a command returns data (denoted in the
register-based command reference by the command name ending with a " ?"
character), that data will be accessed from the Query Response Register
(base+08
) as shown in Figure D-6.
16
)
16
, 0C
, 0E
)
16
16
16
)
16
)
16