OUTPut:VOLTage:AMPLitude
Parameters
Comments
Usage
OUTPut:VOLTage:AMPLitude?
Comments
Usage
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
OUTPut:VOLTage:AMPLitude <
voltage on programmable Strain Bridge Completion SCPs pointed to by
(the HP E1511 Transient Strain SCP, for example).
Parameter
Parameter
Name
Type
<
amplitude
>
numeric (float32)
ch_list
channel list (string)
<
>
•
To turn off excitation voltage (when using external voltage source), program
amplitude
to
.
<
>
0
•
Related Commands:
OUTPut:VOLTage:AMPLitude?
•
*RST Condition:
MIN (0)
OUTP:VOLT:AMPL 5,(@116:119)
OUTPut:VOLTage:AMPLitude? (@<
excitation voltage for the channel specified by
• <
channel
>
must specify a single channel only.
•
Returned Value: Numeric, one of
is float32.
•
Related Commands:
OUTPut:VOLTage:AMPLitude
OUTP:VOLT:AMPL? (@103)
amplitude
ch_list
>,(@<
Range of
Values
MIN | 0 | 1 | 2 | 5 | 10 | MAX
100 - 163
Set excitation voltage for channels 16
through 19.
channel
returns the current setting of
>)
channel
<
0
,
1
,
2
,
5
, or
10
. The C-SCPI type returned
Returns current setting of excitation voltage
for channel 3.
HP E1313/E1413 Command Reference
OUTPut
sets the excitation
>)
<
ch_list
>
Default
Units
none
none
.
>
219