Initiate[:Immediate] - HP E1313A Manual

High speed a/d module
Table of Contents

Advertisement

Usage

INITiate[:IMMediate]

Comments
Usage
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Sending the
command will immediately send the trigger system back to its
ABORt
Trigger Idle state and terminate any scan in progress. If
, however, it will immediately return to the Wait-for-Trigger state. If
ON
TRIGger:SOURce
is
IMMediate
INITiate:CONTinuous OFF
check the " Measuring" bit (bit 4) with
determine when the scan has completed.
Sending
INITiate:IMMediate
(already
) will cause Error -213, " Init ignored" .
INITiated
Related Commands:
ABORt
*RST Condition: Trigger system is in the Trigger Idle state.
INIT:CONT ON
INIT:CONT OFF
changes the trigger system from the Idle state to the

INITiate[:IMMediate]

Wait-for-Trigger state. When triggered, one or more (depending on
) trigger cycles occur and the instrument returns to the Trigger Idle
TRIGger:COUNt
state.
ROUTe:SCAN LIST1
execute.
• INITiate:IMMediate
clears the FIFO and Current Value Table.
If a trigger event is received before the instrument is initiated, Error -211,
" Trigger ignored" error is generated.
If another trigger event is received before the instrument has completed the
current trigger cycle (measurement scan), the Questionable Data Status bit 9 is set
and Error +3012, " Trigger too fast" error is generated.
Sending
INITiate:IMMediate
(already
INITiated
) will cause Error -213,"Init ignored".
Sending the
ABORt
command will immediately send the trigger system to the
Trigger Idle state and terminate any scan in progress.
Related Commands:
ABORt
*RST Condition: Trigger system is in the Idle state.
INIT:IMM
INITiate:IMMediate
also, then the module will resume scanning. Use
to stop after the current scan is complete. You can
STATus:OPERation:CONDition?
while the system is still in the Wait-for-Trigger state
,
TRIGger[:IMMediate]
,
Return to Initiated State after each trigger.
Finish scan (if triggered) and return to
Trigger Idle state AFTER the next trigger
event if already armed.
through
specifies which of four scan lists to
LIST4
while the system is still in the Wait-for-Trigger state
,
TRIGger
Both versions same function.
HP E1313/E1413 Command Reference
INITiate
is
INITiate:CONTinuous
to
TRIGger:SOURce
203