General Form Of The Fifo Data Retrieval Section - HP E1313A Manual

High speed a/d module
Table of Contents

Advertisement

General Form of
the FIFO Data
Retrieval Section
Comments
Chapter 4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Figure 4-1 shows program flow in the FIFO data retrieval section for a single
HP E1313/E1413. The conditions before entering the data retrieval section are:
Signal conditioning is set-up.
Channel EU functions are set.
Scan list is defined.
Instrument initiated (
first decision block in Figure 4-1).
Figure 4-1. General Form of FIFO Data Retrieval
It is not necessary that the instrument be triggered at this time since the
program will loop waiting for readings to appear in the FIFO buffer.
While the instrument is still making measurements (Measuring bit is 1),
the program will stay on the left side of the flow chart.
By executing a FIFO status command, the program will determine
when there are enough readings to transfer.
When there are enough readings (count depends on which status
FIFO command used), flow will fall through to execute the FIFO
" bulk data transfer" command.
After transferring readings, program flow returns to the decision
block which determines if the instrument has completed all
measurements or is still measuring.
If measurements are complete, (Measuring bit is 0), program flow
goes to the right side of the diagram.
Here a FIFO status command is executed to determine if there are
any readings remaining in the FIFO.
If readings are present, the program executes a FIFO data retrieval
command to transfer the remaining readings.
Understanding the HP E1313/E1413
INITiate
sets the Measuring bit checked in the
103