Controlling Keysight B2980
Performing the Limit Test
Performing the Limit Test
Limit test is performed as shown below.
1. Set the required source condition. For details, see "Controlling the Source
Output" on page 1-13.
2. Set the required measurement condition. For details, see "Controlling the
Measurement Function" on page 1-10.
3. Set the composite limit test. See "Setting the Composite Limit Test" on page
1-19.
4. Set the individual limit tests. See "Setting Individual Limit Tests" on page 1-19.
5. Set the trigger condition. See "Setting the Source Output Trigger" on page 1-15
and "Setting the Measurement Trigger" on page 1-11.
6. Enable the channel. See "Enabling the Current/Charge Measurement" on page
1-10.
7. Execute the :INIT command to start limit test.
Note that the DC source channel applies the value set by the
:SOUR:VOLT:TRIG command when it is triggered, even if the channel is
applying the value set by the :SOUR:VOLT command.
8. Read limit test result. See "Reading Limit Test Result" on page 1-19.
For the programming example, see "Pass/Fail Judgement and Math Function" on
page 2-23.
1- 18
Keysight B2980 Programming Guide, Ed ition 1