Language Reference
SENSe Subsystem
Factory Preset
and *RST:
Off
Remarks:
This command is not available when Demod is set to
Off.
Front Panel
Access:
Det/Demod, Demod, FM, Demod View
DETector Subsection
[:SENSe]:
Type of Detection
[:SENSe]:DETector[:FUNCtion] NEGative|POSitive|SAMPle
[:SENSe]:DETector[:FUNCtion]?
Specifies the detection mode.
Negative peak detection displays the lowest sample taken during the
interval being displayed.
Positive peak detection displays the highest sample taken during the
interval being displayed.
Sample detection displays the first sample taken during the interval
being displayed.
Factory Preset
and *RST:
Positive
Front Panel
Access:
Det/Demod, Detector
Det/Demod, Detector, Peak
Det/Demod, Detector, Sample
Det/Demod, Detector, Negative Peak
EBWidth Subsection
[:SENSe]:
Set Emission BW Number of Averages
[:SENSe]:EBWidth:AVERage:COUNt <integer>
[:SENSe]:EBWidth:AVERage:COUNt?
Set the emission bandwidth measurement number of averages.
Remarks:
This command is used for measurements in the
MEASURE
Front Panel
Access:
Meas Setup, Avg Number
5-90