Chapter 8 - Maintenance & Cleaning - Cardinal Health Infant Flow SiPAP Operator's Manual

Table of Contents

Advertisement

Operator's Manual
Chapter 8 - Maintenance & Cleaning
Cleaning
Examine the exterior of the case and the stand for damage and dirt. If necessary
clean the unit and stand. If damage to either is apparent, always seek qualified
Technical advice.
Clean the exterior surfaces of the driver, Transducer Assembly and stand with a mild
soap or liquid disinfectant solution. Do not use cleaning agents that contain
abrasives. Make sure that cleaning agents do not enter the driver through patient
connections or other ports.
CAUTION
Do not immerse any part of the Infant Flow
with gas or steam.
Maintenance
No special maintenance is required by the operator other than that listed below.
There are no operator serviceable parts. The unit must only be maintained and
serviced by an approved service supplier or trained biomedical engineer. Only parts
approved by Cardinal Health may be used in this unit. Refer to the Service Manual or
your Service Supplier for an approved service parts list
WARNING
Oxygen vigorously accelerates combustion. To avoid explosion hazard, do not
use any instrument or other equipment that may have been exposed to oil or
grease contamination.
Calibrate the oxygen analyzer regularly. Calibration of the oxygen analyzer must
be done with the unit off patient.
Regularly check and empty the water trap accessed from the rear panel of the driver
enclosure. Push the button on the bottom of the water trap to release the water into a
suitable waste receptacle.
Disconnect the air and oxygen gas sources when the Infant Flow
use.
675-101-101
Revision H
®
SiPAP™ driver in water or sterilize it
41
®
SiPAP™ is not in