Cleaning

Caution: Kangaroo OMNI™ Enteral Feeding Pump, and power adapters are not designed to be immersed underwater.
Do not immerse Kangaroo OMNI™ Enteral Feeding Pump, or power adapters in water or other cleaning solutions.
Failure to follow the cleaning procedures described herein could result in hazards to users, patients, and clinicians. As
with any A/C powered electrical device, care must be taken to prevent liquid from entering Kangaroo OMNI™ Enteral
Feeding Pump to avoid electrical shock hazard, fire hazard, or damage to electrical components.
Caution: Disconnect Kangaroo OMNI™ Enteral Feeding Pump from A/C power source before thoroughly cleaning.
After thoroughly cleaning, do not connect to an A/C power source until thoroughly dry.
Caution: User should not disassemble the device when cleaning.
If any of the following events occur, do not use Kangaroo OMNI™ Enteral Feeding Pump until it has been properly
cleaned and dried. For assistance, please contact Customer Service.
• Wetting of the power adapter
• Leakage into Kangaroo OMNI™ Enteral Feeding Pump interior

General Cleaning Instructions

Cleaning should be performed as needed. It may also be desirable to define cleaning intervals based on knowledge of
the environment in which the pump is used. Only personnel trained in the cleaning of medical devices should perform
cleaning. A mild, common dish washing liquid detergent should be used for general cleaning. This detergent should
be used with a 20:1 ratio water to detergent mixture.
Inspect the device for any visible soil immediately following usage and/or procedure. If visible soil is present, wipe
down Kangaroo OMNI™ Enteral Feeding Pump with a paper towel moistened with the cleaning solution, removing
all visible soil. Use a brush to remove soil from hard to reach crevices. Users should inspect the device after cleaning
for any visible soil to ensure the device is cleaned between uses. If soil is visibly present, repeat the cleaning steps
before moving to disinfection instructions. After thoroughly cleaning, store the pump and accessories in an area free
of dirt, dust, pet hair, and other contaminants.
Caution: The use of cleaners and disinfectants other than the ones described in the instructions for use may cause
significant damage to Kangaroo OMNI™ Enteral Feeding Pump and may void warranty.

Cleaning Frequency

It is recommended that Kangaroo OMNI™ Enteral Feeding Pump be thoroughly cleaned after each Kangaroo OMNI™
Feeding Set use for a minimum duration of 30 seconds, to prevent bacterial contamination. Further, failure to
thoroughly clean Kangaroo OMNI™ Enteral Feeding Pump can interfere with the function of the Kangaroo OMNI™
Enteral Feeding Pump rotor and sensors, which can increase the occurrence of errors and warning alarms. See below
for methods for cleaning each component. High frequency of cleaning or cleaning with severe chemicals can cause
the plastic housing to crack or break. This can result in electric shock or other risks. Do not clean more than necessary
to meet user/operator needs.
Instructions for Cleaning Kangaroo OMNI™ Enteral Feeding Pump
Pump Housing Cleaning
• Refer to General Cleaning Instructions above before starting.
• Thoroughly clean outside surface with a damp cloth or sponge using a mild detergent. Ensure all cleaning solution
is rinsed or wiped from the device.
• For difficult to clean areas, it is permissible to wash Kangaroo OMNI™ Enteral Feeding Pump under running water.
Avoid submerging Kangaroo OMNI™ Enteral Feeding Pump or washing with high pressure nozzles, which exceeds
the water proof rating of the pump. Kangaroo OMNI™ Enteral Feeding Pump casing has a water proof rating of
"water jet resistant". This allows for washing under running water or wiping with a damp cloth.
• For excessive cleaning needs, avoid use of sharp objects during the cleaning process. Use of a soft bristle brush for
example may be used.
• Users should inspect the device after cleaning for any visible soil to ensure the device is cleaned between uses.
If soil is visibly present, repeat the cleaning steps before moving to disinfection instructions.