If breath rate monitoring is desired, attach the Transducer Interface and the
abdominal sensor. Refer to Chapter 4 for instructions on application of the
Abdominal Respiratory Sensor. Touch the flashing button to confirm.
If an alarm is activated as a result of any of the settings, the button displays a
flashing "X". Alarm conditions must be cleared and all settings must be confirmed
with a green check mark before other screens can be entered.
2.
Alarm set/confirm Screen
Figure 9 – Alarm set/confirm Screen
Touch the NCPAP button or the Alarm Mute / Reset button for 3 seconds to set the
alarms and to move to the next screen. If either button is not touched within 2
minutes, the alarm limits will be set automatically. When the alarm limits have been
set, the screen display changes to the Mode Select Screen with the driver operating
in NCPAP mode.
Note:
Press High Flow meter must be checked through the Start-up screen and be set
during operation. This enables manual breath or back-up apnea breath, where
applicable, to be active.
675-101-101