Performance Specifications

Table 74
Type
Detection type
Performance
Specifications
Light source
Pulse frequency
Excitation
Monochromator
Emission
Monochromator
Reference System:
Timetable programing:
328
Performance Specifications Agilent 1100 Series Fluorescence Detector
Specification
Multi-signal fluorescence
detector with rapid on-line
scanning capabilities and spectral
data analysis
10 fg Anthracene,
Ex=250 nm, Em=400 nm*
RAMAN (H
2
RAMAN (H
2
Ex=350 nm, Em=397 nm, dark
value 450 nm, standard flow cell
time constant=4 seconds
(8 seconds responsetime)
Xenon Flash Lamp,
normal mode20 W, economy
mode 5 W
296 Hz for single signal mode
74 Hz for spectral mode
Range:200 nm - 700 nm and zero-order
Bandwidth:20 nm (fixed)
Monochromator:concave holographic grating, F/1.6, blaze: 300 nm
Range:280 nm - 900 nm and zero-order
Bandwidth:20 nm (fixed)
Monochromator:concave holographic grating, F/1.6, blaze: 400 nm
in-line excitation measurement
up to 4 signal wavelengths,
response time, PMT Gain,
baseline behavior (append, free,
zero), spectral parameters
Comments
see note below this table
O) > 200 (FLF rev A)
see
O) > 400 (FLF rev >A)
Test"
1100 Series FD Reference Manual
"Raman ASTM Signal-to-Noise
on page 119