Agilent Technologies 1100 Series Reference Manual page 151

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2
Carefully remove the rear foam.
Rear
foam
A
A
A
B
Note:
DO NOT unscrew the grounding screw (marked with
X)
1100 Series FD Reference Manual
3
Using the 4 mm hexagonal key, unscrew the five screws,
remove the cover and place it on the bench without
turning it around.
B
4
Using a 2.5 mm hexagonal key, loosen the screw and
remove the condenser. There is a mechanical stop for the
condenser.
Repairing the Fluorescence Detector
A
A
A
B
5
B
153