6 Bring the top of the assembly into the position shown
below and fix the 3 screws.
Next Steps:
•
Reconnect the encoder cable of the EX-monochromator to FLF board.
•
Reinstall the optical unit as described in
•
Enter the replacement in the maintenance logbook.
•
Reset the monochromator settings to default using the diagnose functions.
•
In this case "Wavelength Calibration Lost" is displayed and the instrument stays in Not Ready condition until you
perform a wavelength calibration.
•
Perform a wavelength verification to check the correct positioning of the replaced assemblies, as described in chapter
"Wavelength Verification and Calibration"
1100 Series FD Reference Manual
7 Assure the UV protector cover is installed. Replace the
foam on the optical unit.
"Installing the Optical Unit"
on page 101.
Repairing the Fluorescence Detector
on page 194.
5