Agilent Technologies 1100 Series Reference Manual page 187

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

6 Bring the top of the assembly into the position shown
below and fix the 3 screws.
Next Steps:
Reconnect the encoder cable of the EX-monochromator to FLF board.
Reinstall the optical unit as described in
Enter the replacement in the maintenance logbook.
Reset the monochromator settings to default using the diagnose functions.
In this case "Wavelength Calibration Lost" is displayed and the instrument stays in Not Ready condition until you
perform a wavelength calibration.
Perform a wavelength verification to check the correct positioning of the replaced assemblies, as described in chapter
"Wavelength Verification and Calibration"
1100 Series FD Reference Manual
7 Assure the UV protector cover is installed. Replace the
foam on the optical unit.
"Installing the Optical Unit"
on page 101.
Repairing the Fluorescence Detector
on page 194.
5
189