Performance Specifications - Agilent Technologies 1200 Series User Manual

Fluorescence detector
Hide thumbs Also See for 1200 Series:
Table of Contents

Advertisement

Performance Specifications

Table 3
Type
Detection type
Performance
Specifications
Light source
Pulse frequency
Excitation
Monochromator
Emission
Monochromator
Reference System:
1200 Series FLD User Manual
Performance Specifications Agilent 1200 Series Fluorescence Detector
Specification
Multi-signal fluorescence
detector with rapid on-line
scanning capabilities and spectral
data analysis
10 fg Anthracene,
Ex=250 nm, Em=400 nm*
RAMAN single wavelength
(H
O) > 500 with Ex=350 nm,
2
Em=397 nm, dark value 450 nm,
standard flow cell
time constant=4 seconds
(8 seconds responsetime)
RAMAN dual wavelength (H
> 300 with Ex=350 nm,
Em=397 nm, dark value 450 nm,
standard flow cell
time constant=4 seconds
(8 seconds responsetime)
Xenon Flash Lamp,
normal mode 20 W,
economy mode 5 W
296 Hz for single signal mode
74 Hz for spectral mode
Range:
Bandwidth:
Monochromator:concave holographic grating, F/1.6, blaze: 300 nm
Range:
Bandwidth:
Monochromator:concave holographic grating, F/1.6, blaze: 400 nm
in-line excitation measurement
Site Requirements and Specifications
Comments
see note below this table
see Service Manual for details
see Service Manual for details
O)
2
200 nm - 700 nm and zero-order
20 nm (fixed)
280 nm - 900 nm and zero-order
20 nm (fixed)
2
37