Agilent Technologies 1100 Series Reference Manual
Agilent Technologies 1100 Series Reference Manual

Agilent Technologies 1100 Series Reference Manual

Autosampler thermostat
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Agilent 1100 Series
Autosampler
Thermostat
Reference Manual
s1
Agilent G1330B
Asset # 304029

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 1100 Series

  • Page 1 Agilent G1330B Asset # 304029 Agilent 1100 Series Autosampler Thermostat Reference Manual...
  • Page 2 WA RN ING This guide is valid for A.01.xx revisions of The hardware and/or software described in the Agilent 1100 Series Autosampler Ther- this document are furnished under a license A WARNING notice denotes a mostat software, where xx refers to minor and may be used or copied only in accor- hazard.
  • Page 3 Repairing the ALS Thermostat Module Instructions on repair procedures of the ALS thermostat ALS Thermostat Parts and Materials Lists for identification of parts and materials Specifications Performance specifications of the ALS thermostat 1100 Series AT Reference Manual...
  • Page 4 1100 Series AT Reference Manual...
  • Page 5: Table Of Contents

    Stage 4: Installing the Sample Tray Stage 5: Installing Tray Cover and Front Cover Stage 6: Turning on the Thermostatted Autosampler Stage 7: Update of Control Module Firmware Stage 8: Update of the Agilent ChemStation Software Transporting the Thermostatted Autosampler 1100 Series AT Reference Manual...
  • Page 6 Overview of the Repairing the Thermostat Exchanging the Power Supply Fuses Exchanging Internal Parts Removing the Top Cover and Foam Exchanging Heat Exchanger Fan Exchanging the Heatsink Fans Exchanging the Sensors Exchanging the TCA board Assembling the Main Cover 1100 Series AT Reference Manual...
  • Page 7 ALS Thermostat Parts and Materials Main Assemblies Accessory Kit G1330-68705 Main Assemblies Foam Parts Plastic Parts Heat Exchanger Fan Parts Specifications Performance Specifications Safety Information General Safety Information Radio Interference Sound Emission Agilent Technologies on Internet Index 1100 Series AT Reference Manual...
  • Page 8 1100 Series AT Reference Manual...
  • Page 9 Agilent 1100 Series Autosampler Thermostat Reference Manual Theory of Operation Introduction to the Thermostatted Autosampler ALS Thermostat Operation ALS Thermostat Control and Electronics Temperature Controlled Autosampler Board (TCA) Electrical Connections Agilent Technologies...
  • Page 10: Theory Of Operation

    Introduction to the Thermostatted Autosampler The Agilent 1100 Series autosampler is designed for use with other modules of the Agilent 1100 Series LC system, with the HP 1050 Series, or with other LC systems if adequate remote control inputs and outputs are available. The autosampler is controlled from the Agilent 1100 Series control module or from the Agilent ChemStation for LC systems.
  • Page 11 Theory of Operation Power Supply TCA Board Heat sink fans Heat sink Heat exchanger Condensation leak drain Heat exchanger Fan Figure 1 Overview of the ALS Thermostat 1100 Series AT Reference Manual...
  • Page 12: Als Thermostat Operation

    A fan draws air from the sample tray area and blows it through the channels of the heating/cooling module. The fan speed is determined according to the environmental conditions (e.g., ambient temperature, humidity). In the heating/cooling module the air reaches the temperature of 1100 Series AT Reference Manual...
  • Page 13 The fan speed is controlled according to the temperature of the Peltier elements. During cooling condensation will appear in the heating/cooling module. The condensed water will be guided out of the ALS thermostat. 1100 Series AT Reference Manual...
  • Page 14: Als Thermostat Control And Electronics

    TCA board. + 36 V Line Voltage Bridge Rectifier Primary Board TCA Board ASM Board Temperatur Sensor Peltier Assembly A Peltier Assembly B Peltier Assemblies C + D Figure 3 Block Diagram ALS Thermostat Electronics 1100 Series AT Reference Manual...
  • Page 15: Temperature Controlled Autosampler Board (Tca)

    Peltier elements. The temperature sensor signals and the signals from the hall sensors in the five fans are the input signals for the multiplexer. The output of the multiplexer is connected to the autosampler 1100 Series AT Reference Manual...
  • Page 16 Fan 3 + 4 Regulator 3 Fan 5 + 36 V Auto Range Primary Board Temp. Sensor 1 to 4 to ASM Board Multiplexer Fan Sensor 1 to 5 EPROM Figure 4 Block Diagram TCA Board 1100 Series AT Reference Manual...
  • Page 17: Electrical Connections

    Theory of Operation Electrical Connections Never use cables other than the ones supplied by Agilent Technologies to ensure WA RN ING proper functionality and compliance with safety or EMC regulations. CAN cable to next module Relay contacts Vial number output...
  • Page 18 • The CAN bus is a serial bus with high-speed data transfer. The two connectors for the CAN bus are used for internal Agilent 1100 Series module data transfer and synchronization. • The REMOTE connector may be used in combination with other analytical instruments from Agilent Technologies if you want to use features such as common shut down, prepare, and so on.
  • Page 19 Agilent 1100 Series Autosampler Thermostat Reference Manual Installing the G1330B Thermostat Site Requirements Unpacking the Autosampler Optimizing the Stack Configuration Controller Requirements Installing the G1330B Thermostat Transporting the Thermostatted Autosampler Agilent Technologies...
  • Page 20: Installing The G1330B Thermostat

    Never operate your instrumentation from a power outlet that has no ground WA RN ING connection. Never use a power cord other than the power cord designed for your region. 1100 Series AT Reference Manual...
  • Page 21 Ensure the autosampler is installed in a level position. If a complete Agilent 1100 Series system is to be installed on the bench, make sure that the bench is designed to carry the weight of all the modules. For a...
  • Page 22 Safety standards: IEC, CSA, UL, EN Installation Category II, Pollution Degree 2 ∞ ∞ Using the autosampler at environmental temperatures higher than 50 C (122 WA RN ING may cause the rear panel to become hot. 1100 Series AT Reference Manual...
  • Page 23: Unpacking The Autosampler

    Table 2. Please report missing or damaged parts to your local Agilent Technologies sales and service office. If the thermostatted autosampler was ordered as an upgrade (G1395A) to an existing autosampler, the shipment will also contain the required software upgrades for your Agilent ChemStation.
  • Page 24 Installing the G1330B Thermostat Table 3 ALS Thermostat Accessory Kit Contents G1330-68705 Description Part Number Waste Tube 5063-6527 Waste Tube Assembly G1330-67300 1100 Series AT Reference Manual...
  • Page 25: Optimizing The Stack Configuration

    Figure 6 Figure 7. This configuration optimizes the system flow path, ensuring minimum delay volume. Control Module Detector Column Compartment Solvent Cabinet Autosampler Degasser Pump Thermostat Figure 6 Recommended Stack Configuration (Front View) 1100 Series AT Reference Manual...
  • Page 26 Installing the G1330B Thermostat Analog signal to recorder CAN bus cable Autosampler - Thermostat cable Remote cable Analog signal AC power AC power to recorder GPIB to LC ChemStation Figure 7 Recommended Stack Configuration (Rear View) 1100 Series AT Reference Manual...
  • Page 27: Controller Requirements

    Installing the G1330B Thermostat Controller Requirements Agilent 1100 Series control module and Agilent ChemStation must have the latest revision firmware/software loaded to allow optimum operation with the thermostattable autosamplers. Older revisions might not recognize or have full functionality with the thermostattable autosampler.
  • Page 28: Installing The G1330B Thermostat

    Do not bend the safety flap away from its position, or attempt to remove the safety cover (see Figure 8). Do not attempt to insert or remove a vial from the gripper when the gripper is positioned below the needle. Safety flap Safety cover Figure 8 Safety Flap. 1100 Series AT Reference Manual...
  • Page 29: Stage 1: Preparing The Als Thermostat And Autosampler

    2 Remove the front cover. Press the two snap fasteners on the sides of the cover and move it away. 3 If the ALS thermostat is located on top of another Agilent 1100 Series Module place the waste tube assembly into the top cover of the ALS thermostat and locate the other end in the waste funnel of the module beneath.
  • Page 30 Condensation Leak outlet 5 Install the front cover of the ALS thermostat. 6 Place the autosampler module on top of the ALS thermostat. Make sure that the autosampler is correctly engaged in the ALS thermostat locks. 1100 Series AT Reference Manual...
  • Page 31 This assures that the cold airstream from the ALS thermostat is correctly guided to the tray area of the autosampler. 8 If there is no Agilent 1100 Series module located beneath the ALS thermostat connect the waste tube to the central waste exit of the autosampler and place in a waste vessel.
  • Page 32: Stage 2: Power Cable And Interface Cable Connection

    Agilent 1100 Series control module or ChemStation for LC for more information. In an Agilent 1100 Series system, the individual modules are connected by a CAN cable. NO TE The Agilent 1100 control module can be connected to the CAN bus at any of the modules in the system.
  • Page 33 Installing the G1330B Thermostat 7 Connect additional cables as required (see Figure 13 on page 34). Autosampler - Thermostat Cable Power connector Safety lever Figure 12 Power Connectors and Safety Levers at Rear of thermostatted Autosampler. 1100 Series AT Reference Manual...
  • Page 34 Installing the G1330B Thermostat Relay contacts CAN cable to next module Vial number output Therm-ALS cable Remote RS232C CAN cable to previous module CAN-bus GPIB Thermostat - Autosampler cable Figure 13 Cable Connections. 1100 Series AT Reference Manual...
  • Page 35: Stage 3: Flow Connections

    3 Ensure that the waste tube is positioned inside the leak channel. Pump From pump To metering device To column compartment Plug Waste tube in Leak channel From needle seat To waste Column compartment Figure 14 Hydraulic Connections 1100 Series AT Reference Manual...
  • Page 36: Stage 4: Installing The Sample Tray

    3 Press the front of the sample tray down to secure the tray in the autosampler. If the tray pops out of its position the air channel adapter is not inserted correctly. Figure 15 Installing the Sample Tray 1100 Series AT Reference Manual...
  • Page 37 The vial positions of the right-hand half tray begin at position 101 as follows: Left-hand 40-position tray: 1–40 Left-hand 15-position tray: 1–15 Right-hand 40-position tray: 101–140 Right-hand 15-position tray: 101–115 1100 Series AT Reference Manual...
  • Page 38 Installing the G1330B Thermostat Position 1 Position 101 Figure 16 Numbering of Tray Positions. 1100 Series AT Reference Manual...
  • Page 39: Stage 5: Installing Tray Cover And Front Cover

    Press the stop fastener to secure it in the right side cover of the autosampler. 3 Close the tray cover. Tray cover Autosampler front door Figure 17 Installation of Tray Cover and Front Cover of the Autosampler 1100 Series AT Reference Manual...
  • Page 40: Stage 6: Turning On The Thermostatted Autosampler

    When finished with the update the control module reboots again. 8 Check that the correct firmware was loaded by pressing “System” (F5) - “Records” (F4). 9 Disconnect the control module and remove the PC card by pressing the card-eject button. 1100 Series AT Reference Manual...
  • Page 41: Stage 8: Update Of The Agilent Chemstation Software

    Enter and then HPUPDATE and press Enter only if your Agilent ChemStation is installed in the directory C:\HPCHEM. If your Agilent ChemStation is installed in a different directory, e.g. D:\HPCHEM, you need to type HPUPDATE D:\HPCHEM and press Enter. 1100 Series AT Reference Manual...
  • Page 42 A: and press Enter and then HPUPDATE and press Enter only if your Agilent ChemStation is installed in the directory C:\HPCHEM. If your Agilent ChemStation is installed in a different directory, e.g. D:\ HPCHEM, you need to type HPUPDATE D:\HPCHEM and press Enter. 1100 Series AT Reference Manual...
  • Page 43: Transporting The Thermostatted Autosampler

    Also, ensure the vial tray is secured in place with suitable packaging, otherwise the tray may become loose and damage internal components. 1100 Series AT Reference Manual...
  • Page 44 Installing the G1330B Thermostat 1100 Series AT Reference Manual...
  • Page 45 Agilent 1100 Series Autosampler Thermostat Reference Manual Troubleshooting and Test Functions Overview of the Thermostat’s Indicators and Test Functions Status Indicators ALS Thermostat Error Messages Agilent Technologies...
  • Page 46: Troubleshooting And Test Functions

    This chapter shows only the error messages that are generated for the ALS thermostat. For detailed description of the autosampler error messages see “Autosampler Error Messages” in your corresponding Reference Manual. 1100 Series AT Reference Manual...
  • Page 47: Status Indicators

    Two status indicators are located on the front of the autosampler. The lower left indicates the power supply status, the upper right indicates the thermostatted autosampler status. Status indicator Power supply indicator Figure 18 Location of Status Indicators 1100 Series AT Reference Manual...
  • Page 48: Power Supply Indicator

    Usually, an error condition requires attention (for example, leak, defective internal components). An error condition always interrupts the analysis. 1100 Series AT Reference Manual...
  • Page 49: Als Thermostat Error Messages

    This section describes the meaning of ALS thermostat module error messages, and provides information on probable causes and suggested actions how to recover from error conditions. 1100 Series AT Reference Manual...
  • Page 50: Fan Failed

    • One of the 4 heat-sink fans is defective. • The cooling heating module fan is defective Suggested Actions: See Exchange “"Exchanging the Heatsink Fans" on page 67 or “"Exchanging Heat Exchanger Fan" on page 65. 1100 Series AT Reference Manual...
  • Page 51: Temperature Control Failed (1 - 4)

    / heating. The electronics monitor the current through the Peltier elements. If the Peltier current is out of a specified limit the Peltier element is defective. Probable Causes: • One of the four peltier elements is defective. Suggested Action: contact the Technologies service organization 1100 Series AT Reference Manual...
  • Page 52: Temperature Sensor Failed (1 - 4)

    ALS thermostat the single sensors cannot be replaced separately. Probable Causes: • Temperature sensor is defective Suggested Action: “"Exchanging the Sensors" on page 69”. 1100 Series AT Reference Manual...
  • Page 53: Lost Contact To Als Thermostat

    Suggested Actions: Check cable connection between autosampler and ALS thermostat Connect cable between autosampler and ALS thermostat Exchange ASM board in the autosampler Exchange TCA board in the thermostat contact the Agilent Technologies service organization 1100 Series AT Reference Manual...
  • Page 54: Power Fail For Als Thermostat Module

    • ALS thermostat module not connected to line power • ALS thermostat module fuse(s) defective • ALS Thermostat power supply defective Suggested Actions: Check correct power line connection Check fuse(s) and replace if defective contact the Agilent Technologies service organization 1100 Series AT Reference Manual...
  • Page 55: Temperature Out Of Range

    Make sure that there is enough space on left and right side of the ALS thermostat and that the air channel inlets and oulets are not blocked Make sure that the ambient air temperature is within its specified limits (4°C to 55°C) 1100 Series AT Reference Manual...
  • Page 56: Bad Cooling / Heating Performance

    10 min for the peltier elements to equilibrate Make sure that there is enough space on left and right side of the ALS thermostat and that the air channel inlets and oulets are not blocked contact the Agilent Technologies service organization 1100 Series AT Reference Manual...
  • Page 57 Agilent 1100 Series Autosampler Thermostat Reference Manual Repairing the ALS Thermostat Module Introduction into Repairing the Thermostat Overview of the Repairing the Thermostat Exchanging Internal Parts Agilent Technologies...
  • Page 58: Repairing The Als Thermostat Module

    The thermostat covers should be kept clean. Cleaning should be done with a soft cloth slightly dampened with water or a solution of water and a mild detergent. Do not use an excessively damp cloth that liquid can drip into the autosampler. 1100 Series AT Reference Manual...
  • Page 59: Using The Esd Strap

    2 Unroll the rest of the band and peel the liner from the copper foil at the opposite end. 3 Attach the copper foil to a convenient and exposed electrical ground. Figure 19 Using the ESD Strap 1100 Series AT Reference Manual...
  • Page 60: Overview Of The Repairing The Thermostat

    Repairing the ALS Thermostat Module Overview of the Repairing the Thermostat Figure 20 shows the main assemblies and their locations which can be repaired. Power Supply Board Heat exchanger Heat exchanger Condensation leak drain Figure 20 Main assemblies 1100 Series AT Reference Manual...
  • Page 61: Exchanging The Power Supply Fuses

    5 Remove the fuse from the fuse holder. 6 Insert a new fuse in the fuse holder. 7 Reinsert the fuse holder and fix with the screwdriver. 8 Reinsert the power cables. 9 Switch on the power switch. 1100 Series AT Reference Manual...
  • Page 62: Exchanging Internal Parts

    CAU TI O N ESD kit when handling internal parts. The procedures in this section describe how to exchange internal parts. You must remove the ALS thermostat from the stack in order to open the main cover. 1100 Series AT Reference Manual...
  • Page 63: Removing The Top Cover And Foam

    This will damage the electronics of the modules. 1 Remove the front cover by pressing the both clip fasteners 2 Move the lever towards the power socket. on both sides of the cover. Lever 1100 Series AT Reference Manual...
  • Page 64 4 Unscrew the screws on the top plate and remove the plate top cover (2). by lifting its back first and then sliding to the front. Screws 5 Unplug all wires at the TCA Board and remove them from 6 Remove the top foam. top foam. 1100 Series AT Reference Manual...
  • Page 65: Exchanging Heat Exchanger Fan

    2 Carefully lift the Fan upwards. wire from the Top Foam. 3 Remove the Fan from the Fan Gasket. 4 Insert the new Fan in the Fan Gasket. See arrow at the side of Fan for correct airflow direction. 1100 Series AT Reference Manual...
  • Page 66 Repairing the ALS Thermostat Module 5 Reinstall the fan and route the wire in the appropriate 6 Reconnect the wire to the connector. guide channel. 1100 Series AT Reference Manual...
  • Page 67: Exchanging The Heatsink Fans

    Note the orientation of the fans. 3 Reinsert the Top Foam and route the cables in the guide 4 Reconnect the connectors in the correct order (See label channels in the Top Foam. on top of Power Supply). 1100 Series AT Reference Manual...
  • Page 68 Repairing the ALS Thermostat Module 5 Replace the metal cover and top cover. 1100 Series AT Reference Manual...
  • Page 69: Exchanging The Sensors

    63). Parts required Sensor flexboard, PN G1330-66504 Heatsink Compound, PN 6040-0454 1 Unscrew the six holding screws on top of the hardfoam 2 Carefully lift the upper part of hardfoam encapsulation encapsulation. away. Screws 1100 Series AT Reference Manual...
  • Page 70 5 Insert some Heatsink Compound into the four sensor 6 Install the new sensor flexboard carefully by pressing seats, to assure good contact. each sensor individually into the seats, and then route the flexboard through the sealing adapter channel. 1100 Series AT Reference Manual...
  • Page 71 8 Reinstall the hardfoam encapsulation. seat. 9 Route the cables in the guide channels and reconnect the 10 Reinstall the Top Plate and the Top Cover. connectors in the correct order (See label on Power Supply). 1100 Series AT Reference Manual...
  • Page 72: Exchanging The Tca Board

    "Removing the Top Cover and Foam" procedure on page 63), item 1 to 4. Parts required TCA board, part number G1330-66500 1 Remove all connectors on the TCA Board. 2 Unscrew the connector screws for the Autosampler to Thermostat cable. 1100 Series AT Reference Manual...
  • Page 73 3 By carefully pulling back on the upper part of the TCA 4 Remove the TCA board. board unsnatch it from the holder. 5 Reinsert the new TCA board. 6 Gently push forward on the upper part of the board, until it snatches into place. 1100 Series AT Reference Manual...
  • Page 74 Repairing the ALS Thermostat Module 7 Refit the connector screws and reconnect all connectors. 8 Replace the top foam section, metal cover and top cover. 1100 Series AT Reference Manual...
  • Page 75: Assembling The Main Cover

    1 Place the top part on the bench and insert the left and right 2 Replace the cover. side into the top part. Front 3 Replace the ALS thermostat into the stack and reconnect the cables and capillaries. 4 Turn on the ALS thermostat. 1100 Series AT Reference Manual...
  • Page 76 Repairing the ALS Thermostat Module 1100 Series AT Reference Manual...
  • Page 77 Agilent 1100 Series Autosampler Thermostat Reference Manual ALS Thermostat Parts and Materials Main Assemblies Accessory Kit G1330-68705 Main Assemblies Foam Parts Plastic Parts Heat Exchanger Fan Parts Agilent Technologies...
  • Page 78: Als Thermostat Parts And Materials

    Table 4 Main Assemblies Item Description Part Number Fuse - Power Supply (T2.5A/250V; CSA, UL listed) 2110-0015 Fuse TCA - Board (T3A/250V; CSA, UL listed) 2110-0029 Front Cover 5062-8582 Cable, autosampler - ALS thermostat G1330-81600 1100 Series AT Reference Manual...
  • Page 79: Accessory Kit G1330-68705

    ALS Thermostat Parts and Materials Accessory Kit G1330-68705 Table 5 Accessory Kit Item Description Part Number Waste Tube 5062-2463 Waste Tube Assembly G1330-67300 * Reorder Number (5 m) 1100 Series AT Reference Manual...
  • Page 80: Main Assemblies

    ALS Thermostat Parts and Materials Main Assemblies Figure 22 Main Assemblies Table 6 Main Assemblies Item Description Part Number Heat sink fans 3160-0884 TCA Board G1330-66500 Heat exchanger fan 3160-1079 1100 Series AT Reference Manual...
  • Page 81: Foam Parts

    ALS Thermostat Parts and Materials Foam Parts Figure 23 Foam Parts Table 7 Main Assemblies Item Description Part Number Top Foam G1330-40102 Bottom Foam G1330-40103 1100 Series AT Reference Manual...
  • Page 82: Plastic Parts

    ALS Thermostat Parts and Materials Plastic Parts Figure 24 Plastic Parts Table 8 Main Assemblies Item Description Part Number Cabinet Kit, includes base, top and sides G1330-68703 Front Cover 5062-8582 Leak Pan 5042-1307 1100 Series AT Reference Manual...
  • Page 83: Heat Exchanger Fan Parts

    ALS Thermostat Parts and Materials Heat Exchanger Fan Parts Figure 25 Heat Exchanger Fan Parts Table 9 Main Assemblies Item Description Part Number Fan Gasket G1330-44104 Heat Exchanger Fan 3160-1079 1100 Series AT Reference Manual...
  • Page 84 ALS Thermostat Parts and Materials 1100 Series AT Reference Manual...
  • Page 85: Specifications

    Agilent 1100 Series Autosampler Thermostat Reference Manual Specifications Performance Specifications Agilent Technologies...
  • Page 86: Performance Specifications

    Specifications Performance Specifications Table 10 Performance Specifications Agilent 1100 Series thermostatted autosampler Type Specification Temperature range: setable from 4°C to 40°C in 1° increments 1100 Series AT Reference Manual...
  • Page 87 Agilent 1100 Series Autosampler Thermostat Reference Manual Safety Information General Safety Information Radio Interference Sound Emission Agilent Technologies on Internet Agilent Technologies...
  • Page 88: A Safety Information

    Any adjustment, maintenance, and repair of the opened instrument under voltage should be avoided as much as possible. When inevitable, this should be carried out by a skilled person who is aware of the hazard involved. Do not 1100 Series AT Reference Manual...
  • Page 89 Safety Symbols Symbol Description The apparatus is marked with this symbol when the user should refer to the instruction manual in order to protect the apparatus against damage. Indicates dangerous voltages. Indicates a protected ground terminal. 1100 Series AT Reference Manual...
  • Page 90 A caution alerts you to situations that could cause a possible loss of data. Do not CAU TI O N proceed beyond a caution until you have fully understood and met the indicated conditions. 1100 Series AT Reference Manual...
  • Page 91: Radio Interference

    If test and measurement equipment is operated with equipment unscreened cables and/or used for measurements on open set-ups, the user has to assure that under operating conditions the radio interference limits are still met within the premises. 1100 Series AT Reference Manual...
  • Page 92: Sound Emission

    This product has a sound pressure emission (at the operator position) < 70 dB. • Sound Pressure Lp < 70 dB (A) • At Operator Position • Normal Operation • According to ISO 7779:1988/EN 27779/1991 (Type Test) 1100 Series AT Reference Manual...
  • Page 93: Agilent Technologies On Internet

    Safety Information Agilent Technologies on Internet For the latest information on products and services visit our worldwide web site on the Internet at: http://www.agilent.com/go/chem 1100 Series AT Reference Manual...
  • Page 94 Safety Information 1100 Series AT Reference Manual...
  • Page 95: Index

    23, delay volume, fan, 67, parts and materials, delivery checklist, exchanging, performance specifications, flow path, physical specifications, fuses, 20, power consideration, power cords, power requirements, power supply indicator, half-tray combinations, 1100 Series AT Reference Manual...
  • Page 96 #1, 67, 69, security lever, shipping, 23, simple repairs, site requirements, specifications, 22, status indicator, status indicators, storage, temperature, top cover, transport, transport mechanism, unpacking, unpacking the autosampler, vial racks, vial tray, vials, weight, 1100 Series AT Reference Manual...
  • Page 98 In This Book This manual contains information about the Agilent 1100 Series thermostatted for autosamplers. Information for the individual autosamplers can be obtained from the corresponding autosampler reference manual. This supplemental manual describes the following: • installing the thermostat, • theory of operation, •...

Table of Contents