Repairing the Fluorescence Detector

Removing the Optical Unit

When required
Tools required
Parts required
When working on the optical unit, a clean workbench with ESD protection mat must be
CAU TI O N
available. Otherwise optical components or electronic boards may be damaged.
DO NOT remove the foam from the optical unit unless it is required during a procedure.
NO TE
Otherwise parts can fall out.
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Disconnect capillaries.
Remove detector from stack and place it on the
workbench.
Remove the covers as described in
Covers"
on page 144.
Remove the flow cell.
178
For following repairs: power supply, FLL board, Trigger Pack/FLF board and
PMT.
Screwdriver POZI 1 PT3
Hexagonal key, 4 mm
Optical unit G1315-69002 (exchange assembly, comes with foam) or
individual parts depending on the following procedures
1 Disconnect cables (EM monochromator) and (EX
"Removing the
EM
EX
monochromator) from the detector main board (FLM).
A
A
A
B
1100 Series FD Reference Manual
B