Place the pad with the holder on the stand-offs. The
"TOP" sign should be directed towards the trigger pack.
Refit the screws.
Note:
During the next step no horizontal movement should
be made to the pad.
13
Replace the lamp housing cover and fix the screws.
Carefully insert the rear foam. Reconnect the
monochromator cables to the main board. Left is
excitation, right is emission.
Rear
foam
A
EM
A
A
EX
B
1100 Series FD Reference Manual
12
While pressing the top pad down onto the stand-offs,
tighten the screws. Turn the trigger pack such as its
marker stays in line with the "TOP" sign of the brass
holder.
Next Steps:
Replace the covers as described in "Replacing the Foam
and the Top Cover" on page 149.
Re-install the detector in the stack.
Enter the replacement of the lamp in the maintenance
logbook (this will reset the lamp counter).
Turn the lamp on.
Perform a wavelength verification to check the correct
positioning of the replaced assemblies, as described in
chapter
page 101.
B
Repairing the Fluorescence Detector
"Wavelength Verification and Calibration"
5