Agilent Technologies 1100 Series Reference Manual page 103

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Figure 34
6 Load the method WLEMTEST.M. The FLD will change into the
7 Define in Sample Information the file name and description.
8 Start a run. After the 1 minute run has completed, enter the Data Analysis
1100 Series FD Reference Manual
Emission Spectrum
multi-excitation mode and scan in the range of the expected maximum of
±
350 nm
20 nm.
and evaluate the spectrum. The maximum should be found at
±
350 nm
3 nm.
Troubleshooting and Test Functions
4
105