1 Carefully remove the bottom foam piece by sliding it out
towards the rear.
3 Press down the power switch light pipe to remove it from
the coupler.
Power
switch light
pipe
The repair of the power supply assembly is comprised of exchanging the complete assembly. No serviceable
parts are inside.
1100 Series FD Reference Manual
Repairing the Fluorescence Detector
2 Unscrew the power supply at the rear of the module.
4 Remove the power supply completely.
Re-use the coupler on the new power supply.
Coupler
5
191