Exchanging the Lamp

When required
Tools required
Parts required
Wait until the lamp has cooled down before touching the lamp and/or triggerpack
WA RN ING
assembly.
The instrument should not be operated without Xenon flash lamp. Otherwise the
Triggerpack electronics will be damaged.
When loosening or removing screws take care that they do not fall into the instrument. This
NO TE
may result in a complete removal of other assemblies.
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Remove the flow cell or disconnect the tubings from the
flow cell.
Remove detector from stack and place it on the
workbench.
Remove the covers as described in
Covers"
on page 144.
1100 Series FD Reference Manual
Lamp: if noise or intensity exceeds limits or lamp does not ignite
Screwdriver POZI 1 PT3
Hexagonal screwdriver 2.5 mm, 100 mm long
Hexagonal screwdriver 4mm, 100 mm long
Flash lamp 2140-0600
1
"Removing the
EM
EX
Repairing the Fluorescence Detector
Disconnect the emission and the excitation cables and
from the detector main board.
A
A
A
B
5
B
147