Agilent Technologies 1100 Series Reference Manual page 243

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

The radiation source is a xenon flash-lamp. The 3
continuous spectrum of light from 200 nm to 900 nm. The light output
distribution can be expressed as a percentage in 100 nm intervals, see
Figure
sensitivity requirements. You can economize during automatic operation using
keyboard setpoints, so the lamp flashes during your analysis only. The lamp
can be used until it no longer ignites, but the noise level may increase with
usage.
UV degradation, especially below 250 nm is significantly higher compared to
Visible wavelength range. Generally the "LAMP ON during run" - setting or
using "economy mode" will increase lamp life by a magnitude.
Figure 78
1100 Series FD Reference Manual
78. The lamp can be used for some 1000 hours depending on the
Relative
Intensity
300
400
Lamp Energy Distribution (vendor data)
Introduction to the Fluorescence Detector
500
600
700
Wavelenght [nm]
s flash produces a
µ
800
900
7
245

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents