Agilent Technologies 1100 Series Reference Manual page 223

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Table 43
Connector
01046-60210
START and STOP are connected via diodes to pin 3 of the 3394 connector.
NO TE
Agilent 1100 to 3396A Integrators
Table 44
Connector
03394-60600
Agilent 1100 to 3396 Series II / 3395A Integrators
Use the cable 03394-60600 and cut pin #5 on the integrator side. Otherwise
the integrator prints START; not ready.
1100 Series FD Reference Manual
Identifying Parts and Materials
Pin
Pin
3394
Agilent 1100
13, 15
Pin
Pin
3394
Agilent 1100
9
1 - White
NC
2 - Brown
3
3 - Gray
NC
4 - Blue
NC
5 - Pink
NC
6 - Yellow
5,14
7 - Red
1
8 - Green
NC
9 - Black
13, 15
Signal Name
Active
(TTL)
Not connected
Signal Name
Active
(TTL)
Digital ground
Prepare run
Low
Start
Low
Shut down
Low
Not connected
Power on
High
Ready
High
Stop
Low
Start request
Low
Not connected
6
225