Flash Trigger Lost

This message is displayed when the flash trigger is no longer generated.
Probable Causes
• firmware problem
• defective encoder
Suggested Actions
96
Reboot the detector (power cycle).
Multi Mode Off: exchange the firmware or replace the main board.
Multi Mode Ex: replace EX monochromator assembly.
Multi Mode Em: replace EM monochromator assembly.