Wavelength Calibration Lost - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Troubleshooting and Test Functions

Wavelength Calibration Lost

After exchanging the monochromator assemblies, the calibration factors
should be reset to defaults values (a new FLM board comes with default
values). In this case "Wavelength Calibration Lost" is displayed and the
instrument stays in a Not Ready condition.
Probable Causes
• Reset of monochromator settings after exchange.
• Replacement of FLM board.
Suggested Actions
98
Perform a wavelength calibration.
1100 Series FD Reference Manual