Performance Specifications - Agilent Technologies 1260 Infinity II User Manual

Multi-angle light scattering detector
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

Introduction

Performance Specifications

Table 3
Type
Measurement range size
Measurement range mass
Maximum pressure load
Solvent compatibility
Cell material
Cell volume
Cell heating
Data rates
Detection Range
Laser
Safety Functions
Liquid connections
Status Indicators
Software
Multi-Angle Light Scattering Detector User Manual
Detector Specifications
Description
8 nm – 500 nm (R
10
Da – 10
3
5 bar (500 kPa; 72.5 psi)
Aqueous and organic solvents compatible with the wetted material,
pH range: 2 – 11 (pH value higher 11 is considered damaging)
no strong oxidizing acids
Glass, titanium, and PTFE + 25% carbon
63 µL
10 °C above room temperature and up to 50°C
0 = off
Stability: ± 0.01°C at 35°C cell temperature
Safety switch at 70 °C
No active cell cooling
Maximum sampling rate: 6 Hz
12° - 164° at 20 angles
Wavelength: 660 nm at 10 – 120 mW
Lifetime: 10 000 hours
Leak sensor
Vapor sensor
Coned 10-32 UNF ports for tubing with 1/16 inch outer diameter
LASER LED: (green) ON when the laser is active
ERROR LED: (red)
Agilent WinGPC Software
Da (without Mie correction, depending on ��������/��������)
), depending on sample
g
9
ON if a vapor, leakage, or an external error
occurred
24