Troubleshooting and Test Functions

Using the Built-in Test Chromatogram

This function is available from the Agilent ChemStation and the Control
Module.
The built-in Test Chromatogram can be used to check the signal path from the
detector to the ChemStation and the data analysis or via the analog output to
the integrator or data system. An example is shown in
chromatogram is continuously repeated until a stop is executed either by
means of a stop time or manually.
Figure 52
Built-in Test Chromatogram (default settings)
The peak height is always the same but the area and the retention time depend on the set
NO TE
peakwidth, see examples below.
124
Figure
52. The
1100 Series FD Reference Manual