Data flow for chromatographic output

The data flow is shown below.
Influences
PMT gain
Peakwidth
Responsetime
Figure 96
For multi-channel output, any single flash data point will be processed in a
separate independent filter.
1100 Series FD Reference Manual
PMT Raw Signal
acquisition
position
Dark
subtraction
Divide
PMT/Reference
Diode
Filtering
Channel A, B, C, D
Output
(LU)
Data flow for chromatographic output
Theory of Operation
Reference Diode
Raw Signal
acquisition position
Dark
subtraction
Normalization of
reference signal
Peak Detector
8
Influences
EX wavelength
will be executed
at each flash
assigned for
chromatographic
output
277