Repairing the Fluorescence Detector

Replacing the Mirror Assembly

When required
Tools required
Parts required
DO NOT touch the surface of any optical component during this procedure. This will reduce
NO TE
the reflection efficiency.
One side of the mirror is marked "R" representing the rear side.
NO TE
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Remove the flow cell.
Remove detector from stack and place it on the
workbench.
Remove the covers as described in
Covers"
on page 144.
172
If reflection is too low (contaminated)
Screwdriver POZI 1 PT3
A pair of tweezers
Mirror assembly 1000-1000
1
"Removing the
EM
EX
Disconnect the emission and the excitation cables and
from the detector main board.
A
A
A
B
1100 Series FD Reference Manual