5
Repairing the Fluorescence Detector

Replacing the Foams and Covers

Note:
If a new optical unit has been installed or assemblies within the optical unit have been
replaced, then refer "Next Steps" on
When required
Tools required
Prerequisites
The front cover (across the optical unit) should only be removed when required during a
NO TE
procedure. This will keep dust away from optical components.
1
Carefully insert the rear foam.
196
page
189.
When all repairs have been completed
Screwdriver POZI 1 PT3
The detector is open and other procedures have been carried out
2
Route the cables towards the main board and reconnect
to their connectors.
Rear
foam
EM