Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
Use a flat screw driver to lift the leak base from the base.
7
Replace the leak pan into the cabinet bottom. Press the
leak pan completely down into its recess.
Next Steps:
Replace the covers as described in
Re-install the flow cell and the front panel
Re-install the detector in the stack.
1100 Series FD Reference Manual
6
B
8
B
"Replacing the Foams and Covers"
Repairing the Fluorescence Detector
Remove the leak base completely.
B
Replace the leak sensor using your left hand and
reconnect the connector.
on page 196.
5
167