Forced Cold Start Settings - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Table 69
One start bit and one stop bit are always used (not selectable).
Per default, the module will turn into 19200 baud, 8 data bit with no parity.

Forced Cold Start Settings

Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning switches 1 and 2 to other positions (other than being both
up) will allow for normal operation.
Forced cold start erases all methods and data stored in the non-volatile memory.
CAU TI O N
Exceptions are diagnose and repair log books which will not be erased.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
Table 70
To return to normal operation, set switches back to your GPIB or RS 232
configuration settings.
1100 Series FD Reference Manual
Parity Settings
Switches
Parity
7
8
0
0
No Parity
1
0
Odd Parity
1
1
Even Parity
Forced Cold Start Settings
Mode Select
1
TEST/BOOT
1
2
3
4
1
0
0
Theory of Operation
5
6
7
0
0
0
8
8
1
293