Performance Specifications

Table 3
Type
Detection type
Light source
Wavelength range
Short term noise
(ASTM)
Drift
Linearity
Wavelength accuracy
Band width
Flow cells
Control and data
evaluation
Analog outputs
1200 Series Variable Wavelength Detector User Manual
Performance Specifications Agilent 1200 Series Variable Wavelength Detector
Specification
Double-beam photometer
Deuterium lamp
190–600 nm
-5
± 0.75 × 10
-4
3 × 10
AU/hr at 254 nm
> 2 AU (5%) upper limit
± 1 nm
6.5 nm typical
Standard: 14-µl volume, 10-mm
cell path length and 40 bar
(588 psi) pressure maximum
High pressure: 14-µl volume,
10-mm cell path length and
400 bar (5880 psi) pressure
maximum
Micro: 1-µl volume, 5-mm cell
path length and 40 bar (588 psi)
pressure maximum
Semi-micro: 5-µl volume, 6-mm
cell path length and 40 bar
(588 psi) pressure maximum
Agilent ChemStation for LC
Recorder/integrator: 100 mV or
1 V, output range 0.001 – 2 AU,
one output
Site Requirements and Specifications
Comments
AU at 254 nm
See NOTE on
See NOTE on
See NOTE on
Self-calibration with deuterium lines,
verification with holmium oxide filter
Can be repaired on component level
2
page
26.
page 26
page 26
25