Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2 Carefully remove the rear foam (A). Leave the other foam
on the assembly.
Rear
foam
A
A
A
B
4 Disconnect the optical unit cable (J18) and the EM motor
cable (J20).
Next Steps:
Depending on the replacement you want to perform, proceed to the following procedures.
1100 Series FD Reference Manual
3 Disconnect the leak sensor cable (J3) and unscrew the
grounding screw.
Screw
B
5 Remove the optical unit and place it on the bench (ESD
protection required).
Repairing the Fluorescence Detector
5
J3
B
179