Performance Specifications; Specifications - Agilent Technologies 1260 User Manual

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

2
Site Requirements and Specifications

Performance Specifications

Performance Specifications

Specifications

Table 13
Performance Specifications G1315C/D and G1365C/D
Type
Specification
Detection type
1024-element photodiode array
Light source
Deuterium and tungsten lamps
Data rate
up to 80 Hz (G1315C/G1365C)
up to 20 Hz (G1315D/G1365D)
Wavelength range
190 – 950 nm
Short term noise
< ± 0.7·10
(ASTM) Single and
Multi-Wavelength
Drift
< 0.9·10
Linear absorbance
> 2 AU (5 %) at 265 nm
range
Wavelength accuracy
± 1 nm
Wavelength bunching 1 – 400 nm
Slit width
1, 2, 4 , 8, 16 nm
Diode width
< 1 nm
38
-5
AU at 254 and 750 nm
-3
AU/h at 254 nm
Comments
The UV-lamp is equipped with RFID tag that
holds lamp typical information.
see "Specification Conditions" below
see "Specification Conditions" below
see "Specification Conditions" below
Self-calibration with deuterium lines, verification
with holmium oxide filter
Programmable in steps of 1 nm
Programmable slit
Agilent 1260 Infinity DAD and MWD User Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents