Repairing the Fluorescence Detector

Simple Repairs

On the following pages repairs are described that can be carried out without
opening the main cover.
Table 23
Simple Repairs
Procedure
Flow cell exchange
Flow cell flushing
Leak sensor drying
Leak handling System
replacement
134
Typical Frequency
If application requires a different flow cell type or if
defective.
If flow cell is contaminated
If leak has occurred.
If broken or corroded.
Notes
Complete Assembly
A wavelength calibration check should
be performed after replacement.
If the flow cell is removed and inserted,
then a quick calibration check is
performed. If this fails, you must do a
wavelength recalibration, see
"Wavelength Verification and
Calibration"
on page 101.
Check for leaks.
Check for leaks.