Agilent Technologies 1100 Series Reference Manual page 179

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2 Unscrew screws A and B and disconnect connectors C
and D.
D
4 Remove the FLF board carefully in an upwards motion and
place it on an ESD protected mat.
PMT assembly
Safety actuator
1100 Series FD Reference Manual
A
B
C
FLR board (REF
diode)
Repairing the Fluorescence Detector
3 Disconnect the wires from the connector. Use a flat screw
driver to release the wires.
5 If you need to replace either the board or the PMT
assembly, remove the PMT assembly from its socket.
6 Replace the PMT assembly or the FLF board.
7 Re-install the PMT assembly in the socket.
5
181