Agilent Technologies 1100 Series Reference Manual page 123

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Table 22
Peakwidth = >0.05 min (response time = 1 s fast)
Peakwidth = >0.20 min (response time = 4 s STD)
Procedure using the Agilent ChemStation
1 Load the FLD default parameter
2 Complete the Sample Information (no injection).
3 Type into the command line (or as pre-run command):
4 The logbook shows "Preparing for a simulation".
5 Start the run.
The results are processed like a regular chromatogram.
The test chromatogram is switched off automatically at the end of the run.
NO TE
1100 Series FD Reference Manual
peakwidth to >0.2 minutes (standard),
set stop time to 4.8 minutes.
PRINT SENDMODULE$(LFLD,"SIMU:DFLT").
Troubleshooting and Test Functions
Stop time = 1.2 minutes
Stop time = 4.8 minutes
4
125