Exchanging the multi-channel gradient valve on the quaternary pump
Hide thumbs Also See for 1100 Series:

Advertisement

Quick Links

Exchanging the Multi-Channel Gradient Valve on the G1311A,
This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material.
The information contained herein is intended for use by informed individuals who can and must determine its fitness for their purpose.
Windows
and Windows NT
®
®
a25720c.doc
1100 Series Quaternary Pump
are registered trademarks of Microsoft Corp
http:// www.chem.agilent.com
http:// www.chem.agilent.com
http:// www.chem.agilent.com
http:// www.chem.agilent.com
Page 1 1 1 1 of 6 6 6 6

Advertisement

loading

Summary of Contents for Agilent Technologies 1100 Series

  • Page 1 1100 Series Quaternary Pump This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information contained herein is intended for use by informed individuals who can and must determine its fitness for their purpose.
  • Page 2 Repairing the Pump Exchanging the Multi-Channel Gradient Valve (MCGV) Exchanging the Multi-Channel Gradient Valve (MCGV) N O T E The lifetime of the multi-channel gradient valve can be maintained by regularly flushing the valve, especially when using buffer solutions. If using buffer solutions, flush all channels of the valve with water to prevent precipitation of the buffer.
  • Page 3 Repairing the Pump Exchanging the Multi-Channel Gradient Valve (MCGV) 1 Disconnect the connecting tube and the 2 Press the lower sides of the cover to unclip it. solvent tubes from the MCGV, unclip them Remove the cover. from the tube clips and place them into the solvent cabinet to avoid flow by hydrostatic pressure.
  • Page 4 Repairing the Pump Exchanging the Multi-Channel Gradient Valve (MCGV) 5 Replace the MCGV cover. Reconnect the 6 Reconnect the tube from the active inlet valve waste funnel with the waste tube holder in to the middle position of the MCGV and then the top cover.