Installing The Optical Unit - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
Repairing the Fluorescence Detector

Installing the Optical Unit

When required
Tools required
For safe operation make sure that the ground cable is reconnected.
WA RN ING
Preparations for this procedure:
All previous work has been completed.
Power supply has been installed.
Fan has been installed.
The optical unit has its foam cover attached.
194
When all repairs on the optical unit have been completed
Screwdriver POZI 1 PT3
Hexagonal key, 4 mm
1 Place the optical unit in the instrument. Reconnect the
grounding screw.
Screw
1100 Series FD Reference Manual
J3

Hide quick links:

Advertisement

Table of Contents
loading