4
Troubleshooting and Test Functions
The Wavelength Calibration Procedure
When required
Tools required
Parts required
Steps
1 Preparation of the Glycogen Calibration Sample.
2 Preparation of the Flow Cell.
3 Wavelength Calibration.
4 Verification using
Preparation of the Glycogen Calibration Sample
1 To prepare 10 ml of the calibration solution you have to use 10 mg of the
2 Fill the prepared amount into a suitable bottle/vial.
3 Fill 10 ml of distilled water into the vial and shake.
4 Wait 5 minutes and shake again. After 10 minutes the solution is ready.
110
If application requires, or see
Laboratory balance
Glycogen Calibration Sample, Syringe, needle, sample filter and PEEK fitting
from the Accessory Kit
"Check of Raman Band of Water"
Glycogen sample (a tolerance of
Table 18
on page 101.
on page 104.
±
20% is not critical).
1100 Series FD Reference Manual