Agilent Technologies 1100 Series Reference Manual page 43

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Figure 14
The obtained data are combined to setup the time-table for the excitation
wavelength for best limit of detection and selectivity. The optimized switching
events for this example are summarized in Table 8.
Table 8
Time [min]
0
8.2
19.0
This timetable gives the conditions for optimum detection based on the results
of two chromatographic runs.
1100 Series FD Reference Manual
Optimization of the time-program for the excitation wavelength
Timetable for the analysis of 15 polynuclear aromatic hydrocarbons
Exitation Wavelength [nm]
260
260
260
First Steps with the Fluorescence Detector
Emission Wavelength [nm]
350
420
500
2
45