8
Theory of Operation

Communication Settings for RS-232C Communication

The communication protocol used in this instrument supports only hardware
handshake (CTS/RTS).
Switches 1 in down and 2 in up position define that the RS-232C parameters
will be changed. Once the change has been completed, the instrument must be
powered up again in order to store the values in the non-volatile memory.
Table 66
Use the following tables for selecting the setting which you want to use for
RS-232C communication. The number 0 means that the switch is down and 1
means that the switch is up.
Table 67
Table 68
292
Communication Settings for RS-232C Communication
Mode Select
1
RS-232C
0
Baud Rate Settings
Switches
3
4
5
0
0
0
0
0
1
0
1
0
0
1
1
Data Bit Settings
Switch 6
Data Word Size
0
7 Bit Communication
1
8 Bit Communication
2
3
4
1
Baud rate
Baud Rate
Switches
3
9600
1
1200
1
2400
1
4800
1
5
6
7
Data
Parity
Bits
Baud Rate
4
5
0
0
9600
0
1
14400
1
0
19200
1
1
38400
1100 Series FD Reference Manual