Declaration of Conformity for HOX2 Filter

We herewith inform you that the
holmium oxide glass filter (type Hoya HY-1)
meets the following specification of absorbance maxima positions:
Agilent Technologies guarantees the traceability of the specified absorbance
maxima to a National Institute of Standards & Technology (NIST) holmium
oxide solution standard with a lot-to-lot tolerance of ± 0.3 nm.
The wavelength calibration filter built into the Agilent Technologies UV-visible
detectors
Table 78
Product number
is made of this material and meets these specifications. It is, therefore,
suitable for wavelength calibration of these detectors within the specified
wavelength accuracy of the respective detector over its wavelength range.
November 1, 1995______________
DateQuality Manager
340
360.8 nm – 418.5 nm – 536.4 nm (spectral bandwidth: 1 nm).
79853C
G1314A
Wavelength Accuracy
HP 1050 Series
± 2 nm
Agilent 1100 Series
± 1 nm
1100 Series FD Reference Manual