Agilent Technologies 1100 Series Reference Manual page 193

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

2 Reconnect the leak sensor cable (J3) using your left hand.
4 Insert the rear foam and reconnect cables (EM
monochromator) and (EX monochromator) to the detector
main board (FLM).
Rear
foam
A
EM
A
A
EX
B
1100 Series FD Reference Manual
3 Reconnect the optical unit cable (J18) and the EM motor
cable (J20).
J18
Next Steps:
Replace the covers as described in
Foams and Covers"
Re-install the flow cell and the front panel
Re-install the detector in the stack.
Turn the lamp on.
Perform a wavelength verification to check the correct
positioning of the replaced assemblies, as described in
chapter
page 101.
B
Repairing the Fluorescence Detector
on page 196.
"Wavelength Verification and Calibration"
5
J20
"Replacing the
on
195