2
Site Requirements and Specifications
Performance Specifications
Performance Specifications
Specifications
Performance Specifications G7114A
Table 3
Feature
Detection type
Light source
Number of signals
Maximum data rate
Noise
Drift
Linearity
Wavelength range
Wavelength accuracy
Wavelength precision
Slit width
Time programmable
Agilent InfinityLab LC Series VWD User Manual
Performance Specifications 1260 Infinity II Variable Wavelength Detector (G7114A)
Specification
Double-beam photometer
Deuterium lamp
Single and dual wavelength detection
120 Hz (single wavelength detection)
2.5 Hz (dual wavelength detection)
-5
AU, at 230 nm (single wavelength detection)
<±0.25·10
-5
AU, at 230 nm and 254 nm (dual wavelength detection)
<±0.80·10
-4
<1·10
AU/h, at 230 nm
>2.5 AU upper limit
190 – 600 nm
±1 nm, self-calibration with deuterium lines, verification with holmium oxide
filter
<±0.1 nm
6.5 nm typical over whole wavelength range
Wavelength, polarity, peak width, lamp on/off
30