Replacing The Slits - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
Repairing the Fluorescence Detector

Replacing the Slits

When required
Tools required
Parts required
DO NOT touch the surface of any optical component during this procedure. This will
NO TE
deteriorate the reflection process.
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Remove the flow cell.
Remove detector from stack and place it on the
workbench.
Remove the covers as described in
Covers"
on page 144.
176
If required.
Screwdriver POZI 1 PT3
A pair of tweezers
Slit EX G1321-08103
Slit EM G1321-08102 (also used for PMT)
1 Carefully remove the front foam holding the foam as
"Removing the
shown.
1100 Series FD Reference Manual