Installing the Fluorescence Detector
3 Assemble the column detector capillary from the
4 Assemble the waste tubing from the accessory kit.
accessory kit. One side is already factory-assembled.
Pre-assembled
Note:
The fluorescence detector should be the last module in the flow system. An additional detector should be
installed before the fluorescence detector to prevent any overpressure to the quartz cell (maximum 20 bar) .
When working with detector behind the FLD (on own risk) determine the backpressure of this detector first by
- removing the column and the last detect and measuring system pressure at the application flow rate.
- connecting the last detector (without column and FLD) and measuring the system pressure with flow.
- the difference in measured pressure is due to the back pressure generated by the last detector and is the
pressure seen by the FLD.
1100 Series FD Reference Manual
25