Wavelength Calibration - Agilent Technologies 1100 Series Reference Manual

Fluorescence detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Troubleshooting and Test Functions

Wavelength Calibration

1 From the user interface start the FLD Wavelength Calibration (see
If the wavelength calibration process fails, refer to
NO TE
page 97.
2 If a deviation is displayed, press Adjust and OK. The history table will be
To look at the history table (ChemStation) start a wavelength calibration and abort
NO TE
immediately. No changes are made to the calibration at this time.
Rinse the flow cell with pure water at a minimum of 1.5 ml/min to get rid of the Glycogen
NO TE
from the cell and the capillaries. When organic solvent is sequentially applied (without
rinsing), a blockage of capillaries may occur.
112
Figure
41).
Agilent ChemStation: Diagnosis - Maintenance - FLD Calibration
Control Module: System - Tests - FLD - Calibrate.
updated.
"Wavelength Calibration Failed"
1100 Series FD Reference Manual