Installing the Fluorescence Detector
Installing the Detector
Preparations
Parts required
1 Install the LAN interface board in the detector (if required), see
2 Place the detector in the stack or on the bench in a horizontal position.
3 Ensure the line power switch at the front of the detector is OFF.
Status indicator
green/yellow/red
Line power switch
with green light
Figure 5
4 Connect the power cable to the power connector at the rear of the detector.
5 Connect the CAN cable to other Agilent 1100 Series modules.
6 If a Agilent ChemStation is the controller, connect either
7 Connect the analog cable(s) (optional).
22
Locate bench space
Provide power connections
Unpack the detector
Detector
Power cord, for other cables see below and
ChemStation and/or Control Module G1323A/B.
the Interface Board"
Front View of Detector
•
the GPIB cable to the detector or
•
the LAN connection to the LAN interface board in the detector.
on page 199.
"Cable Overview"
on page 218,
"Replacing
1100 Series FD Reference Manual