5
Repairing the Fluorescence Detector

Replacing the EM-Condenser or Cutoff-Filter

When required
Tools required
Parts required
Preparations for this procedure:
Turn off the detector.
Disconnect the power cable.
Remove the flow cell.
Remove detector from stack and place it on the
workbench.
Remove the covers as described in
Covers"
on page 144.
Remove the flow cell.
168
If light throughput is too low
Screwdriver POZI 1 PT3
Hexagonal key 2.5 mm
Condenser assembly EM 1000-1124
Cutoff filter 295 nm 5062-8530
1
"Removing the
EM
EX
Disconnect the emission and the excitation cables and
from the detector main board.
A
A
A
B